Objective The initiation of cardiopulmonary resuscitation (CPR) can be complicated by the use of protective equipment in contact sports, and the rate of success in resuscitating the patient depends on the time from incident to start of CPR. The aim of our study was to see if (1) previous training, (2) the presence of audiovisual feedback and (3) the presence of football shoulder pads (FSP) affected the quality of chest compressions.
Methods Six basic life support certified athletic training students (BLS-ATS), six basic life support certified emergency medical service personnel (BLS-EMS) and six advanced cardiac life support certified emergency medical service personnel (ACLS-EMS) participated in a crossover manikin study. A quasi-experimental repeated measures design was used to measure the chest compression depth (cm), rate (cpm), depth accuracy (%) and rate accuracy (%) on four different conditions by using feedback and/or FSP. Real CPR Help manufactured by ZOLL (Chelmsford, Massachusetts, USA) was used for the audiovisual feedback. Three participants from each group performed 2 min of chest compressions at baseline first, followed by compressions with FSP, with feedback and with both FSP and feedback (FSP+feedback). The other three participants from each group performed compressions at baseline first, followed by compressions with FSP+feedback, feedback and FSP.
Results CPR performance did not differ between the groups at baseline (median (IQR), BLS-ATS: 5.0 (4.4–6.1) cm, 114(96–131) cpm; BLS-EMS: 5.4 (4.1–6.4) cm, 112(99–131) cpm; ACLS-EMS: 6.4 (5.7–6.7) cm, 138(113–140) cpm; depth p=0.10, rate p=0.37). A statistically significant difference in the percentage of depth accuracy was found with feedback (median (IQR), 13.8 (0.9–49.2)% vs 69.6 (32.3–85.8)%; p=0.0002). The rate accuracy was changed from 17.1 (0–80.7)% without feedback to 59.2 (17.3–74.3)% with feedback (p=0.50). The use of feedback was effective for depth accuracy, especially in the BLS-ATS group, regardless of the presence of FSP (median (IQR), 22.0 (7.3–36.2)% vs 71.3 (35.4–86.5)%; p=0.0002).
Conclusions The use of audiovisual feedback positively affects the quality of the depth of CPR. Both feedback and FSP do not alter the rate measurements. Medically trained personnel are able to deliver the desired depth regardless of the presence of FSP even though shallower chest compressions depth can be seen in CPR with FSP. A feedback device must be introduced into the athletic training settings.
- Chest compression
- High quality CPR
- CPR performance
- Feedback
- Football shoulder pads
- American football
This is an Open Access article distributed in accordance with the Creative Commons Attribution Non Commercial (CC BY-NC 4.0) license, which permits others to distribute, remix, adapt, build upon this work non-commercially, and license their derivative works on different terms, provided the original work is properly cited and the use is non-commercial. See: http://creativecommons.org/licenses/by-nc/4.0/
Statistics from Altmetric.com
Footnotes
Acknowledgements The authors thank the Zoll Medical Cooperation for assistance in using the AED Pro and Real CPR Help.
Contributors ST, WR and SS contributed to the study design. ST carried out the all studies, participated in the sequence alignment and drafted the manuscript. HT helped to provide the intervention equipment. All authors performed the data analysis. WR, SS, HT and RS advised and helped to revise the manuscript. ST prepared the manuscript and all authors read and approved the final manuscript.
Competing interests None declared.
Provenance and peer review Not commissioned; externally peer reviewed.
Correction notice This paper has been amended since it was published Online First. Owing to a scripting error, some of the publisher names in the references were replaced with 'BMJ Publishing Group'. This only affected the full text version, not the PDF. We have since corrected these errors and the correct publishers have been inserted into the references.